ORIGINAL ARTICLE
Year : 2014  |  Volume : 9  |  Issue : 3  |  Page : 296-304

Central and Peripheral Corneal Thickness Measurement in Normal and Keratoconic Eyes Using Three Corneal Pachymeters


Ophthalmic Research Center, Shahid Beheshti University of Medical Sciences, Tehran, Iran

Correspondence Address:
Sepehr Feizi
Department of Ophthalmology, Labbafinejad Medical Center, Boostan 9 St., Pasdaran Ave., Tehran 16666
Iran
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/2008-322X.143356

Rights and Permissions

Purpose: To assess the agreement among ultrasonic pachymetry, the Galilei dual Scheimpflug analyzer, and Orbscan II for central and peripheral (Galilei vs. Orbscan) corneal thickness (CCT and PCT) measurement in normal and keratoconic eyes. Methods: In this prospective study, CCT and PCT were measured in 88 eyes of 88 refractive surgery candidates and 128 eyes of 69 keratoconic patients with ultrasonic pachymetry, the Galilei, and Orbscan II. The readings by the three instruments were compared using one-way analysis of normal variance. Agreement among the three devices was assessed using Pearson and intraclass correlation coefficients, and Bland-Altman plots. The same analyses were employed to evaluate agreement between Galilei and Orbscan II for PCT measurement. Results: In the normal group, mean CCT was 551.0±39.4, 566.9±33.5, and 565.5±40.9 μm measured by ultrasonic pachymetry, the Galilei, and Orbscan II, respectively (P<0.001). The corresponding figures in the keratoconus group were 492.0±61.7, 502.0±42.1, and 470.6±56.9 μm, respectively (P<0.001). Mean PCT was 612.5±35.3 and 640.9±38.0 μm in the normal group (P<0.001) and 567.6±35.2 and 595.1±41.4 μm in the keratoconus group (P<0.001) by the Galilei and Orbscan II, respectively. CCT and PCT measurements obtained by different devices were significantly correlated in both groups. Conclusion: To measure CCT, the Galilei and Orbscan II can be used interchangeably in normal eyes, but not in keratoconic eyes. For PCT, there is a systematic error between measurements obtained by the Galilei and Orbscan II. However, it is possible to change optical pachymeter readings into those obtained by ultrasonic pachymetry using a constant.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed998    
    Printed44    
    Emailed0    
    PDF Downloaded126    
    Comments [Add]    

Recommend this journal